Smith & Nephew BIRMINGHAM HIP Disassembly. & Cleaning Instructions Download Page 1

Disassembly and cleaning instruction

Insert-Ball Head Press for POLARCUP

, OR3O

 and

BIRMINGHAM HIP

◊ 

Dual Mobility Insert

SAP/Art. No. 75017089/21000505

This disassembly and cleaning instruction is designed to supplement the instruction 
“Processing (cleaning, disinfection and sterilization) of instruments” Lit. No. 03389.

Key

1. Stand
2. Base
3. Lengthwise guide
4. Pendulum
5. Buffer
6. Handle

Disassembly

In accordance with the instruction “Processing (cleaning, disinfection and sterilization) of
instruments” Lit. No. 03389.

The following applies in addition to or in place of the instruction:
•  Remove the handle (6), pendulum (4), lengthwise guide (3) and buffer (5) from the stand (1)
•  Remove the base (2) from the stand (1)
•  Remove the lengthwise guide (3) with the buffer (5) from the handle
•  Remove the buffer (5) from the lengthwise guide (3)
•  Rotate the handle (6) from the pendulum (4)
•  Do not remove any of the O-rings!