WD3060 - WD5090
USER MANUAL
26
ALARM ID
DISPLAY MESSAGE
MEANING AND SUGGESTED USER ACTION
ID5
“Probe TL1 disconnected”
Working temperature probe “TL1” is generating an abnormal
signal (probe “open”).
.follow the
DEFAULT PROCEDURE
described above.
ID7
“Probe TA1 disconnected”
Drying temperature probe TA1 is generating an abnormal signal
(probe “open”).
.follow the
DEFAULT PROCEDURE
described above.
ID10
“Probe TCL disconnected”
Control temperature probe TLC is generating an abnormal signal
(probe “open”).
.follow the
DEFAULT PROCEDURE
described above.
ID11
“Lack of cold water”
No cold water during filling phases.
.Check the water supply.
.Check that the intake tap is open.
.Check the water supply pressure.
.Check that the filler hoses are correctly positioned.
ID12
“Lack of warm water”
The same as ID11 for the warm water.
.Check the correct setting of the parameters of the device
(effective presence/absence of warm water).
ID13
“Lack of demi water”
The same as ID11 for the demi water.
.Check the correct setting of the parameters of the device
(effective presence/absence of demi water).
ID17
“Cold water load time exceeded”
Cold water filling time not correct.
.The same checks as for ID11
ID18
“Warm water load time
exceeded”
Warm water filling time not correct.
.The same checks as for ID12
ID19
“Demi water load time exceeded”
Demineralized water filling time not correct.
.The same checks as for ID13
ID20
“Water load system failure”
.The same checks as for ID11, 12, 13
ID22
“Cold water flowmeter failure”
.The same checks as for ID11
ID23
“Insufficient water in chamber”
Chamber water level too low.
.The same checks as for ID11, 12, 13
. Verify that there are no leaks from the machine, highlighted by
the presence of water near the device.
ID24
“NO water in chamber”
.The same checks as for ID23.
ID25
“No Press in Hyd Sys Foam(?)”
Anomaly related to the circulating pump “ML”. Circulating pump
pressure too low. There may be foam in the chamber.
.Check the type of detergent used.