WD3060 - WD5090
USER MANUAL
10
5.1
ACCESSING AND REUSING THE DEVICE AFTER AN INCOMPLETE CYCLE
The instructions relating to the device’s safety in the event of an incomplete operating cycle are provided in
compliance with points 5.4.4.g and 13.1.102 of IEC61010-2-040:2005.
WARNING
If a disinfection cycle is interrupted (by the user or due to an alarm generated by the appliance itself):
take
care when handling the instruments
and the load in general in the washing chamber. The load and the
internal parts of the appliance might be biologically contaminated/infected.
Before handling tools or performing any maintenance: Perform a complete disinfection cycle or, if you
cannot perform a complete Thermo-disinfection cycle, handle tools with caution (using protection devices
required for the handling of infected instruments, eg. gloves, lab coat).
DANGER, HOT SURFACES
The appliance performs a thermal disinfection cycle using water at high temperatures (up to 93°C) and
detergents. If, in the event of a failure, there is water in the chamber when the door is open: avoid contact
with the skin, burning hazard and risk of irritation due to the toxicity of the chemicals used.
Never touch the heating elements inside the chamber.
Contact Smeg authorised technical staff in the event of a failure.
Fig. 1- View of inside of chamber, detail of heating elements.
Do not touch
the heating
elements