background image

Installation & Maintenance Manual
Pressure Sensor for General Fluids
PSE575/PSE576/PSE577

PSE57#-TFT44-A

2 Specifications

2 Specifications (Continued)

Warning

Do not disassemble, modify (including changing the printed
circuit board) or repair.

An injury or failure can result.

Do not operate the product outside of the specifications.

Do not use for flammable or harmful fluids.
Fire, malfunction, or damage to the product can result.
Verify the specifications before use.

Do not operate in an atmosphere containing flammable or
explosive gases.

Fire or an explosion can result.
This product is not designed to be explosion proof.

Do not use the product in a place where static electricity is a
problem.

Otherwise it can cause failure or malfunction of the system.

If using the product in an interlocking circuit:

•Provide a double interlocking system, for example a mechanical

system.

•Check the product regularly for proper operation.
Otherwise malfunction can result, causing an accident.

The following instructions must be followed during maintenance:

•Turn off the power supply.
•Stop the air supply, exhaust the residual pressure and verify that the

air is released before performing maintenance.

Otherwise an injury can result.

Caution

Rated
pressure
range

PSE575

PSE577

0 to 2 MPa

Withstand
pressure

Pressure
spec.

Model

Temperature
characteristics

1 Safety Instructions

This manual contains essential information for the protection of users
and others from possible injury and/or equipment damage.
•Read this manual before using the product, to ensure correct handling,

and read the manuals of related apparatus before use.

•Keep this manual in a safe place for future reference.
•These instructions indicate the level of potential hazard by label of

"Caution", "Warning" or "Danger", followed by important safety
information which must be carefully followed.

•To ensure safety of personnel and equipment the safety instructions in

this manual and the product catalogue must be observed, along with
other relevant safety practices.

CAUTION indicates a hazard with a low level of risk
which, if not avoided, could result in minor or
moderate injury.

Caution

Warning

Danger

WARNING indicates a hazard with a medium level
of risk which, if not avoided, could result in death or
serious injury.

DANGER indicates a hazard with a high level of risk
which, if not avoided, will result in death or serious
injury.

This product is class A equipment that is intended for use in an industrial
environment.
There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted as well as radiated
disturbances.

After maintenance is complete, perform appropriate functional
inspections and leak tests.

Stop operation if the equipment does not function properly or there is a
leakage of fluid.
When leakage occurs from parts other than the piping, the product might
be faulty.
Disconnect the power supply and stop the fluid supply.
Do not apply fluid under leaking conditions.
Safety cannot be assured in the case of unexpected malfunction.

Use within the specified operating pressure.

Otherwise it can cause damage to the Pressure Sensor or inability to
measure correctly.
If fluid is supplied at a pressure exceeding the proof pressure, the ceramic
diaphragm will be damaged which may short circuit the power supply.
Please use a power supply which includes short circuit protection.

±5%F.S. (at 25 

o

C)

5 MPa

0 to 10 MPa

30 MPa

PSE576

0 to 5 MPa

12.5 MPa

Piping specifications

Refer to the operation manual on the SMC website
(URL http://www.smcworld.com) for more information about safety
instructions.

3 Installation

1 Safety Instructions (Continued)

02

R1/4, M5 x 0.8

Materials of parts in contact
with fluid

Port size

Model

Weight

Without lead
wire and M12
connector

With lead wire
and M12
connector

103 g

Piping port: C3604 + electroless nickel plated
Pressure sensor: Al

2

O

3

(aluminum oxide)

Square ring: FKM

191 g

Applicable

PSE57 -02

PSE57 -02-28

Gas or liquid that will not corrode the materials of parts in

contact with fluid

Current
consumption

Fluid

Power supply
voltage

Electrical
spec.

Model

Analogue output
Accuracy (at 25 

o

C)

±2.5%F.S.

10 mA or less

12 to 24 VDC±10% (with 10% max. voltage ripple)

Output type

Analogue
output

Analogue output: 1 to 5 V
Output impedance:

Approx. 1 kΩ

Analogue output: 4 to 20 mA
Maximum load impedance:

500 Ω or less (at 24 VDC)
100 Ω or less (at 12 VDC)

Linearity

±0.5%F.S.

Repeatability

±0.5%F.S. (at 25 

o

C)

Standard

CE, RoHS

Protection

Protected against reverse connection

Withstand
voltage

Enclosure

Environ
mental

500 VAC, 1 minute, Between terminals and case

IP65

Insulation
resistance

100 MΩ or more (at 500 VDC)

Between terminals and case

Operating
temperature
range

Operation: -10 to 60 

o

C, Storage: -20 to 70 

o

C

(No condensation or freezing)

Operating
humidity
range

Operation, Storage: 35 to 85%RH (No condensation)

Cable specifications

Analogue output

AWG23

0.72 mm

Cross linked vinyl chloride

Sheath

Material

Insulator

Conductor

Nominal wire
size

Wire diameter
Material

Outside
diameter

Colours

Outside diameter

ø4

Oil resistant vinyl chloride

1.14 mm

Brown, Blue, Black, White

Model

PSE576

0 to 5 MPa

Rated pressure range

0 to 10 MPa

0

A

0

5 MPa

PSE575

0 to 2 MPa

0

2 MPa

B

10 MPa

5

1 to 5 VDC

1

A

B Pressure

A

na

lo

gu

ou

tp

ut

 [V

]

20

4 to 20 mA

4

A

B Pressure

A

na

lo

gu

ou

tp

ut

 [m

A

]

Wiring

Connector pin numbers

When the lead wire and connector (ZS-37-A or ZS-37-B) designated for
the PSE570 is used, the wire colours will apply as shown in the
diagram.

How to connect the body and the lead wire and connector

•Align the lead wire connector with the connector key groove, and insert

vertically.

•Connection is complete when the knurled part is fully tightened. Check

that the connection is not loose.

Connector pin numbers (on the lead wire)

•ZS-37-A

1
4

2
3

1

Pin No.

2

DC(+)

Content

N.C. 

Brown

Colour

White

3
4

DC(-)
Analogue output

Blue

Black

Knurled part

: The unconnected terminals are used in SMC, so please do not connect them.

Lead wire and M12
connector

3 m

PSE577

•ZS-37-B

2

3

1

4

Reviews: