Selection of switch operation (OUT1)
Select the switch operation required referring to the list of outputs (page
Press the
button to select.
[1_n]: Reverse output
[1_P]: Non-Reverse output
Press the
button.
Selection of switch operation (OUT2)
Select the switch operation required referring to the list of outputs (page
Press the
button to select.
[2_n]: Reverse output
[2_P]: Non-Reverse output
Press the
button.
Reference condition selection
Select standard condition or normal condition for the display units.
Press the
button to select.
[ Anr]: Standard condition.
Flow display which is converted in
atmospheric pressure at 20
C, 65%R.H.
[ nor]: Normal condition.
Flow display which is converted in
atmospheric pressure at 0
C.
: Flow rate in the specification is the value at standard condition.
If the display unit standard is changed from standard to normal, use the conversion formula.
Flow rate at standard condition x 0.927 = Flow rate at normal condition
: When [nor] is selected, an Indicator LED (reference condition) [
] appears on the upper left of the screen.
Press the
button.
Setting of initialize mode is completed.
Return to measurement mode.