HRX-OM- T002
Chapter 5 Display and setting of various functions
HRS090 Series
5.7
Run timer, stop timer function
5-15
5.7 Run timer, stop timer function
5.7.1
Run timer and stop timer function
This function starts or stops operation of the product automatically when the
set time has passed. The time can be set according to the user
’s working
hours. Set the circulating fluid temperature in advance.
[Run timer] is a function to start operation after a set time. [Stop timer] is a
function to stop operation after a set time. It is possible to set both [Run
timer] and [Stop timer]. The set time of both the [Run timer] and [Stop timer]
can be 99.5 hours at maximum, in 0.5 hour units.
[When communication is used]
If the communication mode is DIO REMOTE or SERIAL mode, this function
does not operate. DIO REMOTE and SERIAL mode operation/stop signals
have priority.
●Run timer
・
Run timer starts operation after the set time.
If the thermo-chiller is already operating or the pump is operating
independently, this function does not operate even when the set time
has passed.
Operation can start when the product condition is normal and there is no
alarm generated.
・
The [
] light turns ON when the run timer is set. The [
] light is turned
OFF when the operation is started by the run timer.
The [
] light does not turn OFF while the stop timer is being activated.
・
The run timer setting is reset when the main power supply is cut or a
power failure occurs. Please set it again.
●Stop timer
・
The [
]light turns ON when the stop timer is set. The [
] light turns OFF
when the operation is stopped by the stop timer.
The [
] light does not turn OFF while the run timer is being activated
・
The
stop timer setting is reset when the main power supply is stopped or a
power failure occurs. Please set it again.