SKC Disposable Parallel Particle Impactor Operating Instructions Manual Download Page 1

SKC User-loaded Disposable 

Parallel Particle Impactors (PPI)

The patented

 impaction-based SKC Disposable Parallel Particle 

Impactor (PPI

®

) Samplers are designed to match precisely the 

collection e

  ciency curves for respirable and thoracic dust 

specified by ISO 7708/CEN and adopted by ACGIH, CEN, 
and other occupational hygiene organizations. Performance 
data of the PPI Samplers relative to the ISO 7708/CEN criteria 
was published in the

 Journal of Physics,

 Conference Series 151, 

2009, and the data was made part of the OSHA Docket used 
to develop the OSHA 

fi

 nal silica rule (

fi

 nal rule page 16439). 

The thoracic model meets the requirements of NIOSH Method 5524 
for metalworking 

fl

 uids and compounds with ACGIH

®

 thoracic TLV

®

s.  

Disposable PPI Samplers o

 er the convenience of single use to eliminate 

sampler assembly and cleaning; small size for worker comfort, even under helmets or 
other PPE; and a choice of 

fl

 ow rates for maximum 

fl

 exibility in pump options, sample 

duration, and contaminant concentration. 

These  instructions  are for PPI models that 

are NOT preloaded with 

fi

 lters and supports; these models contain only impaction 

substrates.

Performance Pro

fi

 le

Sampling Rate:

  2 L/min respirable or thoracic and 4 or 8 L/min respirable

Sample Pump:

 • 

Universal 

or 

AirChek

®

 Series for 2 and 4 L/min

     

• Leland Legacy for 8 L/min

Sample Time:

 

Dependent on method used. 

Note:

 SKC tests indicate that a particulate 

mass of up to 6.8 mg on the four impaction substrates would not a

 ect PPI 

performance. This amount is equivalent to sampling for 6 hours at 4 L/min in 
environments where respirable mass concentration is 4.76 mg/m

3

 and equals 

50% of total dust. However, labs have reported to SKC that they prefer no more 
than 2 mg on the 

fi

 lter for analytical reasons. Therefore, SKC recommends that 

you work with your lab to determine optimum sample times for your unique 
sampling conditions.

Sample Media

 37-mm, 

5.0-

μ

m PVC 

fi

 lter, 37-mm, 2.0-

μ

m PTFE 

fi

 lter*

(recommended)

:

  (NIOSH 5524), or 37-mm, 0.8-

μ

m MCE 

fi

 lter 

Use cellulose or stainless steel screen for support

Impaction 

 

Four 3/8-in diameter pre-oiled porous plastic discs

Substrate:

  

(

assembled in all Disposable PPIs

)

Analysis:

  

Gravimetric and/or chemical

Body Material:

   Conductive ABS plastic

Dimensions:

  

 

Height: 

 

4.25 in (10.8 cm) - clip to exhaust

    Diameter:

  

1.8 in (4.6 cm)

  

 Depth:

  

1.2 in (3 cm)

Weight:

  

1.1 oz (31.2 gm)

Shelf-life:

  

18 mos from date of manufacture

†  U.S. Patent No. 7,073,402
* Back pressure on PTFE 

fi

 lters can vary within the same lot.

Operating Instructions

 863 Valley View Road, Eighty Four, PA 15330 USA

Tel: 724-941-9701 • www.skcinc.com

Summary of Contents for Disposable Parallel Particle Impactor

Page 1: ... respirable or thoracic and 4 or 8 L min respirable Sample Pump Universal or AirChek Series for 2 and 4 L min Leland Legacy for 8 L min Sample Time Dependent on method used Note SKC tests indicate that a particulate mass of up to 6 8 mg on the four impaction substrates would not affect PPI performance This amount is equivalent to sampling for 6 hours at 4 L min in environments where respirable mass...

Page 2: ...the inlet plate Particles larger than each impactor s 50 cut point are scrubbed and retained by impaction onto the porous oiled impaction substrate contained in each impactor Smaller particles continue to the standard 37 mm collection filter for analysis PPI Performance Protective end cap Base plate Inlet plate contains 4 impac tors with unique cut points preloaded with 4 impaction substrates and ...

Page 3: ...ps insertfirsta37 mmsupportpad and then a conditioned and weighed 37 mm collection filter if using gravimetric analysis into the base plate 3 Lay base plate on a flat surface align inlet plate with base and press down firmly to achieve an even seal 4 Write sample ID on sampling label Adhere sampling label to bottom of base plate 5 Remove protective end cap from exhaust Technical Tidbits Use forcep...

Page 4: ...onnect the exhaust of the Disposable PPI to the inlet of a sample pump 5 Use flexible tubing to connect the inlet of the calibration adapter to the suction port outlet of a calibrator 6 Calibrate to 2 L min for the 2 L min respirable and thoracic model PPIs 4 L min for the 4 L min respirable model or 8 L min for the 8 L min respirable model Follow the calibration instructions in the pump and calib...

Page 5: ...sed loaded sampler See Media and Sampler Preparation 2 Record sample start time on label 3 Clip Disposable PPI onto a worker s collar or lapel in the breathing zone or in the area to be sampled 4 Clip sample pump at the worker s waist or close to the Disposable PPI 5 Use flexible tubing to attach the Disposable PPI exhaust to the inlet of the sample pump 6 Turn on pump and record pertinent sample ...

Page 6: ...ine Go to http dx doi org enter doi 10 1088 1742 6596 151 1 012060 Trakumas S High flow Personal Respirator Dust Sampler for Increased Sensitivity Poster 261 AIHce 2010 Denver CO Trakumas S High flow Personal Sampler to Monitor Exposure to Respirable Crystalline Silica at New Lower TLV IOHA 2010 8th Conference Book of Abstracts Rome p 59 Trakumas S Salter E High Flow Personal Sampler to Monitor Ex...

Page 7: ...pore size pk 100 225 5 37 PTFE Filters 37 mm 2 0 μm pore size with laminated PTFE support for metalworking fluids NIOSH 5524 pk 50 Contact SKC MCE Filters 37 mm 0 8 μm pore size pk 100 225 5 Filter Supports required for User loaded Disposable PPI sampling Select either cellulose or stainless steel Support Pads cellulose 37 mm pk 100 225 27 Support Pads stainless steel 37 mm wide mesh ea 225 26 Acc...

Page 8: ...8 Form 38055 Rev 1901 www skcinc com ...