Safety Tips
Electrical Leakage Test Point
CAUTION:
If the ground is not properly connected to the electrical leakage test
point a false reading may occur.
In the event the electrical leakage test is done, the proper location for this test is on the main
frame at the head section by the ECM box. Labeling on the main frame will indicate proper
location.
Acceptable Levels of Leakage Current
Allowable limits for leakage current measurements for medical devices depend upon function
and intentional contact with patient. Limits are required for UL 60601-01, IEC 60601-1 and
CSA C22.2 No. 601.1. The maximum allowed leakage readings for the
Bari Rehab Platform 2™
is 300 micro amps.
Electromagnetic Compatibility (EMC)
The Bari Rehab Platform 2™ and SW Rest Secure System™ have been tested for compliance
with the EMC requirements. The guidelines in this section will help to ensure the medical
equipment will meet the requirements of the standard.
WARNING:
Medical equipment should not be used, stacked or located on or
around equipment that may create electromagnetic interferences.
Emissions
The Bari Rehab Platform 2™ and SW Rest Secure System™ been type tested and have passed
the requirements of CISPR 11. Observe the following recommendations to minimize radio
frequency emissions in this section and the Electromagnetic Interference section.
Immunity
The Bari Rehab Platform 2™ and SW Rest Secure System™ have been stringently tested for
susceptibility to electromagnetic radiation over the frequency range 80 MHz to 2.5 GHz. The
test was conducted on this bed and passed the requirements of IEC 61000-4-3.
All pins of connectors have passed ESD testing.