4
7
Safety Tips
Electrical Leakage Test Point
CAUTION:
If the ground is not properly connected to the electrical leakage test
point a false reading may occur.
In the event the electrical leakage test is done, the
proper location for this test is on the main frame by
the center joint on either side of the bed. Labeling
on the main frame will indicate proper location.
Acceptable Levels of Leakage Current
Allowable limits for leakage current measurements for medical devices depend upon function
and intentional contact with patient. Limits are required for UL 60601-01, IEC 60601-1 and
CSA C22.2 No. 601.1. Sizewise electrical products run significantly lower leakage readings
than the maximum allowed. Leakage limits for the Sizewise electrical components are defined
in the electrical test procedures for each product.
Electromagnetic Compatibility (EMC)
The SW Evolution™ and SW Rest Secure System™ have been tested for compliance with the
EMC requirements. The guidelines in this section will help to ensure the medical equipment will
meet the requirements of the standard.
WARNING:
Medical equipment should not be used, stacked or located on or
around equipment that may create electromagnetic interferences.
Emissions
The SW Evolution™ and SW Rest Secure System™ have been type tested and have passed the
requirements of CISPR 11. Observe the following recommendations to minimize radio
frequency emissions in this section and the Electromagnetic Interference section.
Immunity
The SW Evolution™ and SW Rest Secure System™ have been stringently tested for
susceptibility to electromagnetic radiation over the frequency range 80 MHz to 2.5 GHz. The
test was conducted on this bed and passed the requirements of IEC 61000-4-3.
All pins of connectors have passed ESD testing.
List of Cables and Accessories
Replacement parts, such as cables and accessories, must be purchased through Sizewise to
ensure proper compliance requirements.