Quick Start Guide SITA ConSpector

 

Monitoring the bath contamination

 

 

D

o

c.

R

e

v1

 0

9

/2

0

1

7

 

Sampling

 

Take  samples  at  a  well-mixed  and  representative 
sampling  location.  Always  carry  out  measurements 
at  the  same  sample  temperature.  Allow  samples  to 
cool  down  to  room  temperature  before  starting  the 
measurement.  

Checking the device 

Check the device regularly to identify contamination 
on  the  immersion  probe  (light  conductor  and  limit-
er). Use the stand and place the immersion probe in 
the clean and empty stainless steel cup; cover it with 
the  respective  lid  (blank  measurement  set-up).  To 
start the device check select Menu>Checking.  

Measuring the bath contamination 

1.

 

Switch on the SITA ConSpector (press  -key for 2 s).  

2.

 

Choose  the  desired  measuring  profile.  The  active 
measuring  profile  appears  at  the  lower  right  edge 
of the display.  Use the arrow  selection keys 

Liquid

 

to change the profile.

 

 

3.

 

Place  the  stainless  steel  beaker  filled  with  approx. 
80 ml of the sample liquid (approx. 1 cm below the 
rim  of  the  beaker)  on  the  support  surface  of  the 
magnetic stirrer. Place a magnetic stick in the beak-
er and turn the rotary knob to mid position.  

4.

 

To  immerse  the  immersion  probe  into  the  sample 
liquid,  lower  the  device  holder  by  using  the  star 
knob  screw  (see  set-up  in  picture  above).  Immerse 
the  immersion  probe  approximately  half  way  into 
the bath sample. The limiter should rest against one 
side of the beaker.  

5.

 

Use the lid to cover the bath sample. 

measuring set-up SITA ConSpector 

active profile 

immersion 
probe 

limiter 

immersion 
depth 