Quick Start Guide SITA ConSpector
Monitoring the bath contamination
D
o
c.
R
e
v1
0
9
/2
0
1
7
Sampling
Take samples at a well-mixed and representative
sampling location. Always carry out measurements
at the same sample temperature. Allow samples to
cool down to room temperature before starting the
measurement.
Checking the device
Check the device regularly to identify contamination
on the immersion probe (light conductor and limit-
er). Use the stand and place the immersion probe in
the clean and empty stainless steel cup; cover it with
the respective lid (blank measurement set-up). To
start the device check select Menu>Checking.
Measuring the bath contamination
1.
Switch on the SITA ConSpector (press -key for 2 s).
2.
Choose the desired measuring profile. The active
measuring profile appears at the lower right edge
of the display. Use the arrow selection keys
Liquid
to change the profile.
3.
Place the stainless steel beaker filled with approx.
80 ml of the sample liquid (approx. 1 cm below the
rim of the beaker) on the support surface of the
magnetic stirrer. Place a magnetic stick in the beak-
er and turn the rotary knob to mid position.
4.
To immerse the immersion probe into the sample
liquid, lower the device holder by using the star
knob screw (see set-up in picture above). Immerse
the immersion probe approximately half way into
the bath sample. The limiter should rest against one
side of the beaker.
5.
Use the lid to cover the bath sample.
measuring set-up SITA ConSpector
active profile
immersion
probe
limiter
immersion
depth