5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.10 Sanitizing
Operating Instructions Simulation Unit SIM MODULAR
5.10.1.1
Sanitizing on the EasyPad
Displaying the sanitizing operating context
✔
All instruments are in their holders.
1. Press the
Torque control/Purge
fixed key.
The
Purge
operating context is shown on the EasyPad display.
2. Press the favorite key 3 twice.
The
Sanitizing
operating context is displayed.
3. Press the favorite key
2 / Setup
.
Error message: Refill with DENTOSEPT P disinfectant
If the
Refill disinfectant
display appears after the sanitation process is
started, the supply of DENTOSEPT P disinfectant in the reservoir of the
supply module is not sufficient to sanitize the simulation unit. Sanitation
cannot be started with insufficient disinfectant; see "Refilling with
DENTOSEPT P disinfectant".
For operation with a stand-alone water supply, the water is first pumped
out of the reservoir of the supply module.
Refill disinfectant
is then
displayed.
➢ Fill approx. 0.6 liter DENTOSEPT P into the reservoir of the supply
module.
When sufficient DENTOSEPT P has been added, the sanitizing
program will continue.
Inserting instruments into the cleaning adapters
The supply module has integrated adapters for water-carrying
instruments. This allows all instruments to be simultaneously treated with
a high concentration of sanitation solution and then rinsed with water. For
this purpose, sanitizable instruments must be inserted into the adapters
on the supply module.
✔
The "Start" text is shown on the EasyPad display.
1. Remove the SPRAYVIT sleeves from the valve bodies and the
straight and contra-angle handpieces from the water-carrying
instruments.
Purge
3
2
Setup