Error processing
båÖäáëÜ
7.4.3
Rescue program (SiRescue.exe)
Explanation
In case of unexpected network errors, there may be problems with image
transfer from the X-ray component to SIDEXIS.
In this case the system closes the connection after a while (? Time-out
settings) and enters the so-called Rescue state. This means that the image is
not lost, but is kept in the RAM of the X-ray component by a safety mechanism
until it is retrieved by the SiRescue rescue program. A further exposure with
this X-ray box is blocked until this moment.
1804
You may need to install/reinstall the sensor being used.
1806
Connect a sensor or check the sensor’s cable or
connection.
1807
The sensor has been successfully replaced or newly
installed. No further action is necessary.
1811
see 1806
1812
It was necessary to initialize the sensor with default values.
This can lead to a loss of image quality. Perform an
installation for this sensor with the related floppy disk
(check the sensor number).
1814
see 1804
1961
A read error has occurred.
1962
Error in TWAIN configuration. If necessary, reinstall your
TWAIN components (scanner, camera, etc.).
1963
-
1964
see 1962
1965
see 1962
1966
see 1962
1967
see 1962
1968
see 1962
1969
see 1962
1970
see 1962
1971
Unable to access clipboard.
CAUTION
The X-ray component must not be switched off during the rescue state
(fast alternate flashing of the sensor and ready LED).
If the box is switched off, the image is lost and the X-ray component will be
ready for operation after the next power on.