63 03 452 D3352
128
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
Following this double selection and confirmation via the Service
key, the service routine is activated.
11.10.2.3
Select parameters
Select parameter XG 3D / XG 3D ready / GALILEOS
If arrow keys are displayed in the selection fields once the required
service routine has been selected, you can use these arrow keys to
choose between different parameters.
Example
You want to run service routine S017.6 to activate the remote control.
✔
Once you have selected service routine S017.6, the code "00" is pre-
selected for the "Remote control disabled" option.
➢ Touch the + or - arrow keys (A) to select the code 01 (B) for the
"Remote control enabled" option.
Once the selected parameter has been changed (in this case the
code for the activation of the remote control), the Memory key (C)
lights up.