11.2
Adjusting the panoramic X-ray unit
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
132
D 3352.031.03.18.02
Manual adjustment of the PAN symmetry
The manual adjustment procedure is similar to the one for
automatic adjustment. The only difference is that the default
adjustment values automatically determined by SIDEXIS
are overwritten by manually determined adjustment values
in the
PAN
-
SYMMETRY
submenu.
1.
Measure distances
S1, S2
and
S3
with the SIDEXIS
measuring ruler.
IMPORTANT
Measure in the lower area of the needles if possible, since
they may have been bent slightly after repeated use.
Always measure exactly from pin center to pin center.
IMPORTANT
Tip:
To facilitate the measuring procedure, you can color
the image in SIDEXIS (see also SIDEXIS Operator’s Manu
-
al).
1.
S2
S1
S3
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......