7 – 43
Tab 7
7.14 Checking the device leakage current
7.14
7.14
Checking the device leakage current
DANGER
Shock hazard! It is essential that you switch the X-ray unit OFF before
replacing any components!
NOTE
i
A high resistance measuring voltage source at line frequency and a measur-
ing circuit compliant with the requirements of IEC 601 are required.
Complete test units, e.g. the "Bender tester", fulfill these requirements.
If you have not done it already...
z
Switch
OFF
the line voltage at the main switch of the building installation.
z
Disconnect the power cable and the second protective ground wire from
the building installation.
z
Remove the following cover parts (see section 1.14):
– Profile cover (10)