7.8 Checking the diaphragm
Tabs 7
7.8.2
Diaphragm test exposures
The diaphragm test exposures are used for a quick check of the system
adjustment. The
TEST
EXPOSURE
menu is started without a service pass-
word. Diaphragm test exposures can thus be performed by the user as well.
Menu: Test exposure
The
TEST
EXPOSURE
menu is started from
SIDEXIS XG
:
EXTRAS
‡
CONSTANCY
TEST
‡
XCXP
‡
SELECT
X
-
RAY
DEVICE
‡
SERVICE
EXPO
-
SURE
‡
DIAPHRAGM
TEST
EXPOSURE
NOTE
i
When you open the
TEST
EXPOSURE
menu, the unit switches from the user
mode to the PC service mode logged by the PC.
In the
PC service mode
, the control options that are available on the Easypad
are determined by SIDEXIS and the currently selected service routine.
General control of the unit (as in the user mode) is not possible in this mode.
You can switch between the
PAN
-
DIAPHRAGM
,
PAN
-
SYMMETRY
,
CEPH
-
FIXED
POINT
OF
ROTATION
and
TSA
-
SENSOR
menus by clicking the
corresponding
tab
with the mouse. To quit the
TEST
EXPOSURE
menu, click
CANCEL
.
NOTE
i
The Ceph - Fixed point of rotation and TSA - Sensor tabs are not present in the
XG 3/3 PPE system version.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...