båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 139
Tab 5
5.19 Service routine S034
5.19
S034.5: Display/calibrate center position of ceph scan secondary diaphragm axis
5.19.2
S034: Test step 5
Selecting test step 5
1.
Select test step 5 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
After test step 5 is selected, the current offset value for the center position of
the ceph scan
secondary diaphragm axis
is displayed in selection field 1.
Switching the test mode on
2.
Press the
T key
to switch the test mode on.
NOTE
i
In the test mode, travel to the center position of the ceph scan secondary dia-
phragm axis is initially performed without offsets. The T key (XG
Plus
) or the
LED above the T key (XG 5) lights up and displays the test mode.
1
2
3
+
-
S034
5
n
(
1
2
3
S034_______5____
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
+
-
(
n
(
o
(
o
S34
(
n
1
2
3
1500
n
(
1
2
3
____1500________
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
(
n
S34
5
R
T
T
T
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...