5.16 Service routine S021 (not for XG 3 / 3 PPE)
5.16
S021.3: Function test of TSA sensor adjustment
5.16.2
S021: Test step 3
Selecting test step 1
1.
Select test step 3 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
An inactive progress indicator in selection field 1 and the message FFFF
in selection field 2 signal that the system is ready for initialization of the di-
aphragm axis.
1
2
3
+
-
S021
3
n
(
1
2
3
S021_______3_____
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
+
-
(
n
(
n
(
o
(
o
T
1
2
3
n
(
1
2
3
________________
XG
Plus
:
Easypad touchscreen
XG 5:
Multipad
S21
3