Tab 4 4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed)
4.
4
4.4.2
Ceph test phantom
Accessory: Ceph test phantom
For the adjustment of the
ceph primary diaphragm
and the
ceph main X-
ray beam direction
, you must insert test phantom
B
in the
sensor slot on
the panoramic X-ray unit
.
For the adjustment of the
ceph fixed point of rotation
, you must remove the
test phantom from the sensor slot on the panoramic X-ray unit.
B