64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
55
Sirona Dental Systems GmbH
5 Installation: X-ray unit
Installation Manual ORTHOPHOS SL
5.4 Installing the floor stand (option 2)
båÖäáëÜ
5.4
Installing the floor stand (option 2)
In case of mounting on weight-bearing
wood structures
Use the enclosed wood screws and washers from the mounting kit for
mounting the unit on weight-bearing wood structures.
NOTICE
Reduced image quality
Even when the unit is installed using the floor stand it must be secured
with the upper wall holder.
NOTICE
Reduced image quality
Note the wall and ground conditions. If a carpet is present at the unit
installation site, it must be removed.
Each wall plug must support a tensile force of 700 N.
The wall construction must be suitable for the installation of the unit.
IMPORTANT
The unit must remain on the pallet until the floor stand has been fully
installed. Only then may the unit be installed. For enhanced
representation, some of the following drawings are shown in the
standing state.
IMPORTANT
The floor stand version is 30 mm (1 3/16") higher than the standard
version.
2x 8x80
2x Ø 8,4