13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.2.3
Sensor calibration
1. In the structure tree, under
"3D"
, click on the
"Sensor calibration"
element (S010.64).
2. Establish [ → 209] receptivity.
3. Create an exposure (85 kV / 22 mA s) [ → 209].
A message is displayed in the action window to indicate that the
exposure is complete and the status indicator in front of the
"Sensor calibration"
element in the structure tree responds
accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
4. Continue with the next stage of the calibration procedure [ → 266].