64 95 142 D3632
194
D3632
.
031.01.02
.
02 03.2016
13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.1 General information on unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1.1
Displays and help messages during adjustment and
calibration
The most frequent help and status messages during calibration are listed
below.
Help messages
H3 01: Move unit to starting position, press the R key.
H3 07: Remove the occlusal bite block.
H3 21: Close the door
H4 02: Plug the sensor into the ceph slot (sensor missing).
H4 03: SIDEXIS 4 is not ready for exposure, make unit ready for
exposure.
H4 04: Plug the sensor into the ceph slot (incorrect sensor plugged in).
Status messages
●
"Ready for exposure"
●
"Exposure not possible"
●
"Please wait"
●
"Ready for exposure in XXs"
●
"X RAY Active!"
●
"Exposure is performed"
If error message E1 11 20 is displayed on the control panel and/or the
remote control during the calibration process, this does not necessarily
indicate an equipment error. This error message only indicates that the
adjustment or calibration data of the unit is incomplete at this point.
Acknowledge the error message with the R key, if applicable, and
continue the adjustment or calibration procedure.
For assistance with other help messages or error messages displayed
during the adjustment or calibration process, please refer to the section
of these instructions entitled “Messages”.
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......