64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
159
Sirona Dental Systems GmbH
11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.4.9
S018: Test step 10
Setting the PWM values for the height adjustment creep speed
In user mode, check whether a correction is required for the speed for
upward and downward movement (creep speed). Here the bite block
must be deflected in such a way that the creep speed is effective in the
relevant direction (observe green arrows). The PWM index values are set
correctly if:
● the drive runs at approximately the same speed in both directions
● the drive spontaneously starts without jolting and concurrent to key
actuation
● the travel speed does not exceed 5 mm/s.
* factory settings
Symbol on the control
panel
Status
Meaning
Patient symbol key 1 (G) visible
Occlusal bite block
connected
not visible
Occlusal bite block not
connected
Selection field
Parameters
Range of values
1
U (upward travel)
1U – 15U* – 35U
2
D (downward travel)
1D – 20D* – 35D