11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.3.8
S017: Test step 15
* Factory setting
Activate/deactivate the acoustic signal for end of exposure
1. Call service routine S017.15.
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting.
4. Exit the service routine.
Selection field
Code
Function
1
00
Acoustic signal indicating the end of the
exposure is disabled
01
Acoustic signal indicating the end of the
exposure is enabled*