64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
135
Sirona Dental Systems GmbH
11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.2.3
Select parameters
If arrow keys are displayed in the selection fields once the required
service routine has been selected, you can use these arrow keys to
choose between different parameters.
Example
You want to run service routine S017.6 to activate the remote control.
✔
Once you have selected service routine S017.6, the code "01" is pre-
selected for the "Remote control enabled" option.
➢ Press the + or - arrow keys (A) to select code 00 (B) for the "Remote
control disabled" option.
Once the selected parameter has been changed (in this case the
code for disabling remote control), the Memory key (C) lights up.
11.12.2.4
Saving parameters
Once one or a number of parameters have been selected via a service
routine, the current selection must be saved so that it is applied in the unit.
Example
You want to run service routine S017.6 to save the selected option
"Remote control disabled".
✔
The Memory key (C) lights up.
1. Touch the Memory key (C).
The R key (D) lights up.
2. Touch the R key (D).
The selected setting is saved to non-volatile memory.
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......