11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.2.2.3
Service routines with security access
A security code is required for accessing service routines involving
functions such as radiation release or editing of configuration data or
stored values. This procedure prevents the inadvertent selection or
activation of these service routines.
To select a service routine or test step with security access, proceed as
follows:
1. Select the service routine or the test step, and confirm your selection
with the Service key [ → 133].
After you have confirmed your selection, a "0" appears in
selection field 2.
2. Confirm security access by once again selecting the number of the
main routine (2 in the example) with the arrow keys in selection field
2 (C) and press the Service key (B) to confirm your selection.
Following this double selection and confirmation via the Service
key, the service routine is activated.