64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
13
Sirona Dental Systems GmbH
2 Safety instructions
Installation Manual ORTHOPHOS SL
2.4 Condensation
båÖäáëÜ
2.4
Condensation
Safety information for condensation: Service engineer
Extreme fluctuations of temperature may cause condensation inside the
unit. Do not switch the unit on before it has reached normal room
temperature. See also Technical Data.
2.5
Qualifications of service personnel
Installation and startup may be carried out only by personnel specifically
authorized by Sirona.
2.6
Switching the unit on
Safety information for switching on the unit: Service engineer
Due to the risk of injury caused by malfunction, no person may be
positioned in the unit when it is switched on.
2.7
Radiation protection
Safety information for radiation protection: Service engineer
The valid radiation protection regulations and measures must be
observed. The statutory radiation protection equipment must be used.
During an exposure, the service engineer should move as far away from
the X-ray tube assembly as the coiled cable of the manual release
permits.
With the exception of the service engineer, no other persons are allowed
to stay in the room during an exposure.
In case of malfunctions, cancel the exposure immediately by letting go of
the exposure release button.
Information about radiation protection for Canada
2.8
Laser light localizer
Safety information for light localizer: Service engineer
The system incorporates Class 1 laser products.
A minimum distance of 10 cm (4") is required between the eye and the
laser. Do not stare into the beam.
Do not use the system with any other lasers, and do not make any
changes to settings or processes that are not described in these
operating instructions. This may lead to a dangerous exposure to
radiation.
NOTICE
3D imaging should not be used for screening examinations. 3D imaging
examinations must be clinically warranted and each exam must be
justified by demonstrating that the benefits outweigh the risks.
NOTICE
Where it is likely that evaluation of soft tissues will be required as part of
the patient's radiological assessment, the appropriate imaging should
be conventional medical CT or MR, rather than 3D imaging using Cone
Beam technology.
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......