63 98 536 D 3439
46
D 3439
.
201.06.10
.
02 10.2017
6 Operation
Dentsply Sirona
6.5 Manual block clamp
Operating Instructions CEREC MC XL Premium Package
6.5
Manual block clamp
Store the block clamp tool in the corresponding holder.
You can attach the holder to a suitable location with the adhesive pad.
Clean and degrease the contact surface beforehand.
Blocks with 10 mm diameter block holders
1. Place the block (A) directly into the block fixing.
2. Clamp the ceramic block securely with the ball pressure screw (B).
Use the block clamp tool with torque wrench for this purpose.
The block is pressed laterally against the contact surface of the
block fixing and simultaneously pulled in axially. The plate of the
block holder thus rests on the block fixing.
NOTICE
Wear of the ball pressure screw
The high clamping forces cause wear of the ball pressure screw.
➢ Replace the ball pressure screw every 500 clamping operations.
NOTICE
Fasten the block tightly
If the block is not sufficiently tightened, it can lead to unsuitable results
and ceramic breakages.
➢ Securely tighten the block with the block clamping tool with the
torque wrench, until you can hear a clicking sound.
➢ Check to make sure that the block is seated very firmly.
Summary of Contents for CEREC MC XL Premium Package
Page 77: ......