1 Caring for the CEREC Bluecam and prismatic tube
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
båÖäáëÜ
1
Caring for the CEREC Bluecam and
prismatic tube
For acquisition in the patient's mouth, always use the CEREC camera
support. This is supplied as a non-sterile part. Disinfect the camera
support prior to application using a disinfectant listed as suitable for the
acquisition unit in the operating instructions.
Grafik_groß
Description
The Bluecam is a very sensitive optical device and must therefore be
handled with the utmost care. Protect the front lens and the prism against
scratching and clean them with a lint-free cloth and ethanol (commercially
available cleaning alcohol).
Removing the prismatic tube
A
Press detent to release
D
Camera support
6 pcs., Order No. 5945360
B
Prismatic tube
E
Protective cap
C
Prism
F
Front lens
C
A
F
E
D
B
D