background image

418472 / A /

1. Removal of the USB flash drive
for data retrieval

Using a 24mm (6–point socket) and ratchet

a

Remove the

USB

cap.

b

Remove the gasket protecting the USB flash
drive.

c

Remove the USB flash drive.

d

During short-term storage the

“red”

cap may

be used in place of the USB flash drive.

Note

Do not remove the

Communication

end-cap.

2. Transducer end-cap removal for
battery replacement

From a suitable workbench

• Remove the nylon retainer cord

(!)

and pull the

end-cap clear of the housing.

Note

Open vent valve

(A)

to release internal pressure.

The safety valve

(B)

indicates overpressure.

3. Battery and O-ring replacement

Using the integrated handle

a

Remove the battery and replace with Lithium
or (optional)NiMH rechargeable batteries.

b

Check and replace O-rings as needed at

(A)

,

(B)

, and

(C)

.

Wideband Autonomous

Transceiver

Setting up the WBAT

Reviews: