© SIKA • Ba_TP17-TPM-Multi_en.doc • 01/2014
14.2 Characteristics series TP 17 …
Series
Characteristics
TP
17 165
TP
17 166
TP
17 200
TP
17 450
TP
17 650
Calibrator
Display range
-50…165 °C -50…165 °C -60…200 °C 0…450 °C
0…650 °C
Reference temperature setting
range
-35…165 °C -35…165 °C -55…200 °C T
R
…450 °C T
R
…650 °C
Tolerance
0,4 °C
0,4 °C
0,4 °C
0,6 °C
0,8 °C
Control stability
0,1 °C
Electrical characteristics
Power supply:
- 90…240 VAC +/- 10 %
50/60 Hz
-/-
-/-
- 230 VAC –15 % / +10 %
-/-
-/-
-/-
50/60 Hz
- 115 VAC –20 % / +10 %
(optional)
-/- -/- -/- -/-
50/60
Hz
Power consumption
~400 VA
~600 VA
~2000 VA
~1000 VA
Fuse:
- 90…240 VAC
6,3 A slow
-/-
-/-
- 230 VAC
-/-
-/-
-/-
10,0 A slow 6,3 A slow
- 115 VAC (optional)
-/-
-/-
-/-
-/-
10,0 A slow
Process variables
Test specimen holder:
- Bore
28 mm
60 mm
28 mm
60 mm
28 mm
- Depth
150 mm
Housing dimensions:
- Width
~210 mm
~147 mm
- Height
~380+50 mm
~330+68 mm
- Depth
~300 mm
~269 mm
Weight
~11,4 kg
~12,1 kg
~14,6 kg
~7,5 kg
Sleeves for smaller
(in 0,5 mm steps)
1,5…25 mm 1,5…55 mm 1,5…25 mm 1,5…55 mm 1,5…25 mm