Operating the Calibrator / Micro Calibration Bath
7.3.1 Automatic control
For certain tasks it can be advantageous to switch off the control, e.g. to carry out settings at
the calibrator / micro calibration bath.
Switching off automatic control:
Press
the
P
key when in calibration mode for approx 5 sec., the main menu opens.
The last selected function appears on the upper
display.
LED SET flashes on the lower display.
Press
the
or
key until
OFF
appears.
Press
the
P
key to confirm.
An alternating display of the current reference
temperature and
OFF
appears on the upper dis-
play.
The current set temperature appears on the lower
display
TAKE NOTICE OF:
The controller is now switched off and the reference temperature will continuously
change and adjust to the room temperature without having to be further regulated.
Switching on the automatic control:
The control is switched off if the following display ap-
pears:
An alternating display of the current reference tem-
perature and
OFF
appears on the upper display.
The current set temperature appears on the lower dis-
play.
Switch the control back on by
pressing
the
P
key for approx. 5 sec., the main
menu opens.
OFF
appears on the upper display.
LED SET flashes on the lower display.
Confirm switching on the controller by pressing the
arrow key until the desired operating mode is dis-
played and confirm this with the
P
key.
PLEASE NOTE:
The control has been reactivated. The calibrator / micro calibration bath is in calibration
mode and the set temperature is targeted.
Menu control OFF
Control OFF setting display
Control OFF setting display
OFF display