SIKA Simulator 
UC mAV 
Calibrator for mA & V 
 

Instruction Manual 

version V01 

 

 

 

 

 
 
 

SIKA Dr. Siebert & Kühn GmbH & Co. KG 
Struthweg 7-9 
34260 Kaufungen 
Germany 

 
Tel   

+49-05605-803-0 

Fax   

+49-5605-803-54 

eMail  

[email protected] 

web   

www.SIKA.net 

 
 
 
 
 
 

Summary of Contents for Simulator UC mAV

Page 1: ...lator UC mAV Calibrator for mA V Instruction Manual version V01 SIKA Dr Siebert K hn GmbH Co KG Struthweg 7 9 34260 Kaufungen Germany Tel 49 05605 803 0 Fax 49 5605 803 54 eMail info SIKA net web www...

Page 2: ...the right to bill the purchaser for the costs of importing repair or replacement parts if the product purchased in a given country has been sent to another country for repair The warranty commitment o...

Page 3: ...aged It is nevertheless wise to make a quick check for any damage that might have been incurred in transport If such is the case immediately make the usual reservations with carrier The following stan...

Page 4: ...ltage DC measurement 15 C 1 2 Current DC measurement 16 C 1 3 Auxiliary functions at current measurement 18 C 1 3 1 Passive or active mode 18 C 1 3 2 Scale 18 C 1 3 3 HART 19 C 1 4 Voltage or current...

Page 5: ...of increments single or cyclic ramp signals Its utilisation is facilitated by a series of improvements Fast access to all the functions Intuitive user interface Graphic display of 160x160 pixels Conne...

Page 6: ...C to 50 C relative humidity 20 to 80 non condensing Extreme operating envelope 15 C to 55 C relative humidity 10 to 80 70 at 55 C Extreme storage and transport envelope 30 C to 60 C without AA batter...

Page 7: ...ected to the measuring circuits the terminals can be dangerous therefore do not place your hands near a terminal whether used or not Never exceed the protection limit values indicated in the specifica...

Page 8: ...e the accompanying documents Earth Conforms with the European Union directives 60V CAT II Pol 2 Category II Pollution 2 Maximum voltage rating with respect to earth 60V A 4 Maintenance The instrument...

Page 9: ...s an error could damage the instrument The polarity is indicated inside the compartment cover The following figure shows how to open the battery compartment and the direction of installation of each b...

Page 10: ...surement and transmission while the other two are for current measurement and transmission and the continuity test Function key F1 Function key F2 Validation key VAL ON OFF key CLEAR key HOLD key Navi...

Page 11: ...n the display is divided into seven horizontal fields The 1st field indicates the mode of operation Measurement transmission or continuity The 2nd field indicates the date time and battery charge The...

Page 12: ...a 24V source to the sensor Active mode the instrument delivers no voltage source Hart mode Caution The current source cannot deliver the current demanded unlooping probable Caution The voltage source...

Page 13: ...ignal conflicts the instrument then enters measurement mode B 1 6 Operating modes There are 5 main operating modes Voltage measurement Current measurement Voltage transmission Current transmission Con...

Page 14: ...ent of the range 500 A up to 24 mA 3 2 mA up to 24 mA 500 A up to 25 mA Mode Continuous Increments Ramp Predefined valve FS linear or quadratic scale Continuous Increments Ramp Predefined valve FS lin...

Page 15: ...firm Note The measurement mode is the mode selected by default The type of function voltage or current is selected using key F1 configuration menu Using the navigation key and go onto the function fie...

Page 16: ...the last min max reset command This command is accessed through key F2 Using the navigation key go onto the min max reset field Press VAL to confirm C 1 2 Current DC measurement Measurement or transm...

Page 17: ...Idc using the navigation keys and Press VAL to confirm Using key F1 go onto the range field and define the range Go into the menu by pressing F2 Select the range using the navigation keys 0 20mA 4 20m...

Page 18: ...ess ranges only Hart compatibility for process ranges only After selecting the function and Rrnge the auxiliary functions described above must be defined C 1 3 1 Passive or active mode Using key F1 go...

Page 19: ...t web www SIKA net The scaling curve for the 0 20mA range is as follows 0 5 10 15 20 0 2 4 6 8 10 12 14 16 18 20 x mA y mA The scaling curve for the 4 20mA range is as follows 0 5 10 15 20 0 2 4 6 8 1...

Page 20: ...onfirming Transmission mode the type of generation must be defined Continuous manual or incremental editing Incremental in steps Single ramp only one ramp signal transmitted Cyclic ramp C 1 4 1 Manual...

Page 21: ...the CONFIGURATION RAMP menu using the navigation key The transmitted voltage can be manually decremented from the maximum programmed value using the navigation key C 1 4 4 Single ramp editing Press ke...

Page 22: ...avigation key The transmitted voltage can be manually decremented from the maximum programmed value using the navigation key The Hold key allows generation to be stopped and restarted Ramp generation...

Page 23: ...gation keys and to increment or decrement the setpoint value C 1 5 Ramp configuration The CONFIGURATION RAMP menu is used for the generation of incremental single ramp cyclic ramp signals in both volt...

Page 24: ...given in seconds and is limited to a maximum of 1000s The CONFIGURATION RAMP menu is accessed using key F2 Use key F2 to move from one field to the next The value is entered using the navigation keys...

Page 25: ...is limited to a maximum of 1000s c Delay The delay corresponds to the timeout that can be programmed between the moment the start of transmission key is pressed and the actual start of generation It...

Page 26: ...ode is OFF and depending on the type of quantity transmitted voltage or current transmission b Low Level Duration Rise High Level Duration Fall and Delay The low level duration rise high level duratio...

Page 27: ...VAL to save the parameters Press CLEAR to quit the menu without saving C 1 6 Continuity function CONTINUITY mode is a MEASUREMENT mode applied to the current channel I I terminals It indicates whether...

Page 28: ...confirm Select the Contrast field using the navigation keys and then confirm Increase or reduce the contrast using the navigation keys and D 2 Adjusting the date and time The date and time can be adj...

Page 29: ...menu This menu is accessed using key F1 Select the Setup field using the navigation keys and then confirm Select the Preferences field using the navigation keys and then confirm Select the Resolution...

Page 30: ...ting the language The interface language can be selected to German English French Spain or Italian in the CONFIGURATION SETUP PREFERENCES menu This menu is accessed using key F1 Select the Setup field...

Page 31: ...strument whose performance is at least as good at that used for the preceding verification Or return the instrument to the address indicated below for verification and calibration Technical specificat...

Page 32: ...on Remarks Continuity Open Closed Considered closed at Z 0 1K open at Z 1k E 2 Transmission function Maximum rated voltage in common mode 60 VDC or VAC E 2 1 Voltage DC transmission Range Measurement...

Page 33: ...info SIKA net web www SIKA net E 3 Power supply Autonomy The SIKA UC mAV is designed to function either with four 1 5 V AA batteries or with a 4 8 V battery pack The following autonomies are given for...