Before storage, we recommend that you consider the following points:
Clean the hand pump and the accessories.
Turn the fine adjustment valve in a clockwise direction until
the thread is no longer visible (
illustration).
Open the pressure relief valve.
Important storage information!
•
Do not store under pressure:
Only store the hand pump with the pressure relief valve open. This ensures that no pres-
sure can be built up by unintentional pumping movements.
•
Openings facing up:
Ensure that the caps of the supply container and the refill bottle are properly closed and
facing up when these items are stored.
8
Disassembly and disposal
CAUTION! Risk of injury!
Never remove the device from a measurement set up in operation.
Make sure that the measurement set up is shut down professionally.
Before disassembly:
Prior to disassembly, ensure that
the measurement set up is switched off and is in a safe and de-energised state.
the measurement set up is depressurised and has cooled down.
Disassembly:
Remove the reference gauge and the test item.
Drain the hand pump if there is still medium in the reservoir.
Disposal hand pump:
IMPORTANT! No household waste!
The hand pump consists of various different materials. It should not be disposed of with
household waste.
Take the hand pump to your local recycling plant
or
send the hand pump back to your supplier or to SIKA.
Disposal hydraulic oil:
IMPORTANT!
Dispose of the hydraulic oil in accordance with the Technical Safety Data Sheet.