Servicing
Instruction Manual VisGuard 2 Extractive
14265E/1
29
8.4
Replacing the suction fan filter cartridge KZTN3
Larger particles are collected by the suction fan filter. In order to prevent clogging, the filter
cartridge must be replaced periodically. The interval depends on the dust load in the sample.
As a rule of thumb, a load of 15 mE/m means the filter has to be replaced after 12 months.
If the load is lower, the interval increases accordingly.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the four knurled screws and filter co-
ver as shown in drawing
KZTN3-MB
.
2.
Remove the filter cartridge and replace it.
3.
Fasten the cover in place with the four knurled
screws.
8.5
Cleaning the sample lines
Particles from the sample can accumulate on the inner walls of the sample lines, which can
distort the measuring results. Therefore, sample lines of more than 50 meters in length must
be cleaned periodically with compressed air.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Uncover the ends of the sample lines.
2.
Clean the lines with compressed air.