Settings
Instruction Manual OilGuard 2 W
14809E/2
53
8.4
Setting the outputs (digital)
Outputs 1/2 are located on the AQ2Basi print near the customer connection terminals and
are designated as relays 1 and 2.
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Inp./outputs
button.
If the desired menu does not
appear, press the arrow at the bot-
tom right.
4.
Press the
Outputs
button.
5.
Select the
O1 Inactive .. O… Inactive
output.
6.
Activate the outputs (multiple selection
possible).
Activated outputs are highlighted
green.
Invert
: Inverts the outputs.
Prio fault
Fault
Warning
Service
Adjustment
Limit 1 .. n
The other buttons named
MR-
Out…
are for automatic measuring
range switching
Reference Man-
ual.
7.
Press the
Meas
button.
The instrument is back in measuring
operation.