Instruction Manual OilGuard 2 Ex
Servicing
78
13003E/3
9.7
Contamination check on the free-fall measuring cell
KPFLJ VA
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove both the sample feed and return from
the measuring cell and let the cell run dry.
2.
Loosen the locking screws (arrows) and remove
the cover on the measuring cell.
3.
Loosen the locking screws (circles) and remove
the inlet pipe (A).
4.
Remove the protective glasses (arrows) from
the holder and clean them.
Use alcohol and a cotton cloth for
cleaning.