Instruction Manual OilGuard 2 Ex
Servicing
70
13003E/3
9.4
Contamination check on the measuring cell KPFL 30
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove both the sample feed and return from
the measuring cell and let the cell run dry.
2.
Loosen the locking screws (arrows) and remove
the measuring cell by tipping it slightly.
3.
Loosen the Allen bolt (circle) and open the
clamp.
4.
Remove the measuring cell cover (A).
5.
Clean the interior of the measuring cell (B) and
the measuring cell windows with a clean,
grease-free cloth.
Use soapy water in the event of stub-
born soiling.