Instruction Manual ColorPlus Ex
10210E/5
7
1.4
Purpose and conformity
Use of the photometer for purposes other than that for which it was designed
can produce incorrect measuring results, possibly with consequential damage
to the process and damage to the photometer itself.
Installation
and operation of the control unit and any supplementary
components in areas at risk of explosion is prohibited (DANGER OF FATAL
INJURY)!
If a control unit SIREL Ex is used to read the manufacturer's documentation
11044DEF advance.
The photometer and its periphery are designed for measuring the absorption of
liquids and gases in areas at risk of explosion designated Zone 1, (Ex d IIC
T3/T4/T5/T6 Ga/Gb).
The photometer complies with the following standards for electrical equipment
and for areas at risk of explosion:
EN 60079-0:2009
General requirements
EN 60079-1:2007
Instrument protection with pressure-proof
encapsulation “d”
EN 60079-26:2007
Operating materials with instrument
protection level (EPL) Ga
EN 60079-28:2007
Protection of equipment and transmission
systems which work with optical emissions
Directive 94/9/EC
Tabelle 1: Standards
The photometer possesses the following certifications:
EC-Type Examination Certificate BVS 10 ATEX E 102 X
BVS 17 ATEX ZQS/E422
IECEx BVS 13.0058X
The current rules of engineering practice were observed in its design and
manufacture. They meet the usual directives with regard to safety and the
obligation to exercise due care.
The photometer and the control unit comply with the electromagnetic
compatibility (EMC) requirements applicable within the European Union (EU)
and with the Low Voltage Directives (LVD) and carry the CE mark.
Use in accordance
with design purpose