Servicing
Instruction Manual AquaMaster
12745E/4
89
8.2.2
Checking the flow rate and cleanliness
The following procedure describes how the flow rate is checked:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the optics unit of the photometer
and fasten onto the docking station.
Section 8.2.1
2.
Check the flow rate of the sample according
to the technical data. Section 2.5
If the flow rate is insufficient,
please refer to Section 8.2.3.
3.
Check the water jet.
Normal jet without bubbles, reflections (e.g.
of the shutter) are visible in the water jet.
Small air bubbles are difficult or impos-
sible to see and may occur in the normal jet!
incorrect
Uneven water
jet after water
stop
This can
occur if air is
drawn into the
hose at the
AquaScat 2 in-
put after a wa-
ter stop and the
instrument is
then operated
again.
incorrect
Too little water
incorrect
Too much water
incorrect
With large air
bubbles