Mounting and installation
Instruction Manual AquaMaster
12745E/4
35
4.8
Mount sensors before commissioning
CAUTION!
Damage to the sensors due to improper handling.
pH sensors and Redox/ORP sensors must be carefully handled. pH sensors have a sensitive
glass membrane; redox sensors have a very fine platinum wire at the measuring tip. These
sensors can be damaged by carelessly touching the measuring tip, by improper cleaning.
Do not let pH and Redox/ORP sensors dry out. If they are not going to be used for a while,
put the tips in a storage solution (e.g. 3 molar solution of potassium chloride).
Touch the electrodes and measuring tips of the pH and Redox/ORP sensors only when
absolutely necessary.
Use only cleaning agents in accordance with Section 8.1.1.3.
Oxygen sensors and Conductivity sensors are mechanically somewhat more robust.
These sensors should nevertheless be handled with sufficient care.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Close measuring cell block inlet (X) and open
measuring cell block outlet (Y).
2.
Slightly lift measuring cell block cover and fold
down.
3.
Swing the locking device away from measuring
cell block.
4.
If a pH or Redox/ORP sensor is present, fill the
measuring cell block halfway with water. This
protects the sensor against drying out.