Refrigerated Centrifuge Sigma 3-30KS
6 Using the centrifuge
Version 07/2012, Rev. 1.22 of 02/06/2021
• sb
55 / 93
Translation of the original operating manual, part no. 0701302
→
6.3.2.11 Changing the contrast
To change the contrast:
• Press and hold the stop button and turn the function knob one notch to
the left. A dialog box will be displayed once the stop button is released.
• Adjust the contrast of the centrifuge display and confirm the change.
Fig. 26: Dialog box for changing the contrast
→
6.3.3
Program mode
A program contains all of the data that are required for a centrifuge run.
Certain sedimentation results can be repeated under identical conditions.
Programs can be loaded, executed, edited, and deleted when the
centrifuge is at a standstill.
A maximum of 60 programs can be stored under the numbers 1-60. The
program "RAPID_TEMP" does not occupy any storage location and cannot
be deleted. It is used to bring the centrifuge to a specific temperature
without any vessels.
"--" means that the values that are currently set are not a stored program.
The programs can be protected against unauthorised use, modification, or
deletion with the aid of a code (see chapter 6.3.2.7 - "Parameters menu").
Stored programs are listed in the "Process library" menu
.