Refrigerated Centrifuge Sigma 3-18KS
6 Using the centrifuge
Version 08/2012, Rev. 1.15 of 15/09/2017
• sb
27 / 89
Translation of the original operating manual, part no. 0701102
Pos : 102 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060 Betrieb====================== =================== @ 27\ mod_1405319198895_68. doc x @ 192423 @ 1 @ 1
6
Using the centrifuge
Pos : 103 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0010 Erst e Inbetriebnahme @ 27\ mod_1405319200130_68. doc x @ 192437 @ 2 @ 1
6.1
Initial start-up
DANGER
• Before the initial start-up, please ensure that your centrifuge is properly
set up and installed (see chapter 5 - "Set-up and connection").
Pos : 104 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 105 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020 Ei nsc halt en (auß er 6- 16HS) @ 27\ mod_1405319201340_68. doc x @ 192451 @ 2 @ 1
6.2
Switching the centrifuge on
• Press the mains power switch.
The display then illuminates. The centrifuge is ready for operation.
Pos : 106 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 107 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0010 Ö ff nen und Sc hließ en des Dec kels - 1 Sc hloss @ 27\ mod_1405319202508_68.doc x @ 192465 @ 3 @ 1
6.2.1
Opening and closing the lid
The lid can be opened if the centrifuge is at a standstill and if the lid key is
illuminated.
• Press the lid key in order to open the lid.
The centrifuge cannot be started if the lid is opened.
• To close, press with both hands slightly on the lid until the electrical lock
is locked.
WARNING
Do not place your fingers between the lid and the housing when closing
the lid. Risk of crushing!
Pos : 108 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 109 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0019 Ei ns etz en von Rotor en und Zubehör ( auß er 8K)-------------------------------- @ 27\ mod_1405319204492_68.doc x @ 192493 @ 3 @ 1
6.2.2
Installation of rotors and accessories
Pos : 110 /100 Sig ma/110 BA Z entrif ugen Sig ma (Proj ekt e)/ 3-18KS-3- 18KH S/ 060 Betrieb/060-0020-0020- 0010 Ei nsetz en des R ot ors 3-18KS-KH S @ 29\mod_1405321818213_68. doc x @ 197757 @ 4 @ 1
6.2.2.1
Installation of the rotor
• Open the centrifuge lid by pressing the lid key.
• Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
• Lower the rotor with its central bore straight down onto the motor shaft.
• Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 7.5 Nm. In doing so, hold the rotor at its outer rim.
WARNING
Once a day or after 20 cycles, the rotor tie-down screw must be loosened
by some turns, and the rotor must be lifted and fastened again. This
ensures a proper connection between the rotor and the motor shaft.