7 Malfunctions and error correction
44 / 79
Version 03/2014, Rev. 2.11 of 11/08/2017
• sb
Translation of the original operating manual, part no. 0700902
Pos : 159 /100 Sig ma/110 BA Z entrif ugen Sig ma (Proj ekt e)/ 3-16L/070 St örungen und F ehl ersuche/ 070- 0010- 0010 N ot entriegel ung 3-16L @ 29\ mod_1405321641164_68.doc x @ 197281 @ 3 @ 1
7.1.1
Emergency lid release
In the event of a power failure, it is possible to manually open the
centrifuge lid.
• Switch off the mains power switch and disconnect the power cord from
the socket.
• Remove the stopper (see figure, item 1) from the opening at the side
panel, e.g. with a screwdriver.
Fig. 26: Position of the opening for the emergency lid release
• The stopper is connected with a string. Pull the string in order to release
the lid.
• Then, reinsert the stopper into the opening.
WARNING
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.