• sb
Translation from the original operating manual, part no. 0700602
Pos : 134 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020-0080 R ot orausw ahl - ohne Tem p @ 28\mod_1405320534301_68. docx @ 194551 @ 4 @ 1
→
6.3.2.8
Rotor selection
This field shows the rotor that is currently being used.
The centrifuge is equipped with an automatic rotor identification system.
After the installation of the rotor, the rotor ID will be displayed.
It is also possible to manually select the rotor.
NOTE
The rotor selection can only be changed when the centrifuge is at a
standstill.
• Select the "Rotor" field and confirm the selection. The indication "set"
flashes in the activated status.
• Turn the right function knob until the desired rotor number is selected.
• Press the right function knob to confirm the input. The selected rotor will
be adopted.
Fig. 16: Rotor selection
NOTE
Attention
– special case!!!
If the rotors 12072 or 12107 are not preselected, rotor 12073 will be
automatically displayed for rotor 12072, and rotor 12133 for rotor 12107.
This setting does not cause any problems for the centrifugation since the
suggested rotors have the same maximum speed.
Pos : 135 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 136 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ Spinc ontrol Univ ersal /062- 0030-0020-0090 Softs tart- und Sof tst op-F unktionen @ 28\m od_1405320535209_68. docx @ 194565 @ 4 @ 1
6.3.2.9
Softstart and softstop function
The softstart function is used to extend the acceleration time, whereas the
softstop function is used to extend the deceleration time. The current
combination is shown on the display.
• Press the stop key repeatedly until the desired combination is
displayed. The possible combinations are called cyclically.
– If "soft start" is displayed, only the softstart function is activated.
– If "soft start" and "soft stop" are displayed, the softstart and the
softstop function are activated.
– If "soft stop" is displayed, only the softstop function is activated.
– If "soft start" is displayed and the "soft stop" indication flashes, the
softstart function and the brakeless deceleration are activated.
– If the "soft stop" indication flashes, the brakeless deceleration is
activated.
– If no indication is displayed, the standard operating mode is
restored.
Pos : 137 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1