Laboratory Centrifuge Sigma 2-16P
6 Using the centrifuge
Version 04/2007, Rev. 2.23 of 28/04/2022
• sb
35 / 76
Translation from the original operating manual, part no. 0700602
NOTE
Pay attention to the marking of the centrifuge (see the illustration below)!
Safety indications on the centrifuge must be kept readable at all times. If
necessary, they must be replaced.
Fig. 8: Safety indication on the centrifuge: Loading of a swing-out rotor
Pos : 111 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 112 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ 062- 0020-0020-0040 Adapt er @ 28\m od_1405320421969_68. docx @ 194103 @ 4 @ 1
→
6.2.2.5
Adapters
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
• Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
• If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.
Pos : 113 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 114 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ 062- 0020-0020-0050 Gefäß e ( nur 2- 16P) @ 28\m od_1405320422786_68.docx @ 194117 @ 4 @ 1
→
6.2.2.6
Vessels
• Load the vessels outside of the centrifuge. Liquids in the buckets or
multiple carriers cause corrosion.
• Fill the vessels carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
• Always fill the tubes up to their useful volume (= the volume that is
stated for the tube).
• After the centrifugation, remove the vessels carefully in order to prevent
the samples from mixing.
• Follow the safety instructions and hazard warnings (see chapter 3 -
Maximum speed for tubes
Some tubes, such as centrifuge glass tubes, microtubes, culture tubes,
fluoropolymer tubes and especially high-volume tubes can be used in our
rotors, buckets, and adapters at higher speeds than their breaking limit.
NOTE
When using glass vessels, the maximum value of 4,000 x g must not be
exceeded (except special high-strength glass tubes; please refer to the
information provided by the manufacturer).