Laboratory Centrifuge Sigma 2-16P
6 Using the centrifuge
34 / 76
Version 04/2007, Rev. 2.23 of 28/04/2022
• sb
Translation from the original operating manual, part no. 0700602
Pos : 110 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0030 Eins et zen v on Z ubehör 4plätzig 2-16P, 2-16KL, 3- 16L, 3- 16KL, 3-18KS @ 63\mod_1554803016350_68. doc x @ 683660 @ 4 @ 1
→
6.2.2.4
Installation of accessories
• Only use vessels that are suitable for the rotor.
• In the case of swing-out rotors, all of the compartments must be
equipped with buckets.
• Always load opposite compartments of the rotors with the same
accessories and fill to avoid imbalance.
Centrifugation with vessels of various sizes
Working with vessel of various sizes is possible. In this case, however, it is
very important that the rotationally symmetrical inserts are identical.
correct
incorrect
Fig. 6: Permissible and impermissible loading of a swing-out rotor with vessels of various
sizes (example illustration)
Centrifugation with low capacity
• Install the sample vessels in a rotationally symmetrical manner so that
the buckets and their suspensions are loaded evenly.
• It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 7: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)