Laboratory Centrifuge Sigma 2-16P
6 Using the centrifuge
Version 04/2007, Rev. 2.23 of 28/04/2022
• sb
31 / 76
Translation from the original operating manual, part no. 0700602
Pos : 103 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/062 Betrieb_Spi nc ontr ol Eas y+Univ ersal/ 062- 0020-0020 Eins etz en v on Rotor en und Z ubehör -------------------------------- @ 27\mod_1405320420206_68. doc x @ 194075 @ 3 @ 1
→
6.2.2
Installation of rotors and accessories
Pos : 104 /100 Sigma/110 BA Z entrif ugen Sigma (Proj ekt e)/ 2-16P/062 Betri eb/062-0020-0020- 0010 Ei nsetzen ei nes Rotors 2- 16P @ 29\m od_1405321530898_68.docx @ 197001 @ 4 @ 1
→
6.2.2.1
Installation of a rotor
• Open the centrifuge lid by pressing the lid key.
• Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
• Lower the rotor with its central bore straight down onto the motor shaft.
• Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 3 Nm. In doing so, hold the rotor at its outer rim.
• Follow the safety instructions and hazard warnings (see chapter 3 -
WARNING
Once a day or after 20 cycles, the rotor tie-down screw must be loosened
by some turns, and the rotor must be lifted and fastened again. This
ensures a proper connection between the rotor and the motor shaft.
CAUTION
The rotor tie-down screw must only be tightened when a rotor is installed
in order to prevent the collet chuck from being expanded.
CAUTION
When using rotors for microtiter plate formats:
Ensure that the plate holders are inserted together with the plates into the
buckets.
NOTE
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft.
Removal of a rotor
• Loosen the rotor tie-down screw by turning it anti-clockwise and remove
the rotor.