Blood bank centrifuge Sigma 8KBS
3 Safety
28 / 125
Version 05/2021, Rev. 1.2 of 28/04/2022
• sb
Translation of the original operating manual, part no. 0703502
→
3.7
Safety devices
→
3.7.1
Lid lock device
The centrifuge can only be started when the lid is properly closed. The
electrical lock must be locked. The lid can only be opened when the rotor
has stopped. If the lid is opened by way of the emergency release system
during operation, the centrifuge will immediately switch off and decelerate
brakeless. If the lid is open, the drive is completely separated from the
mains power supply, i.e. the centrifuge cannot be started (see chapter 7.1.1
- "Emergency lid release").
→
3.7.2
Standstill monitoring system
Opening of the centrifuge lid is only possible if the rotor is at a standstill.
This standstill is checked by the microprocessor.
→
3.7.3
System check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. Errors are detected with extreme sensitivity and
displayed as error messages in a dialog box (see chapter 7.2 - "Table of
error codes").
3.7.4
Earth conductor check
An earth conductor check can be carried out by authorised and specialised
personnel using a suitable measuring instrument. Please contact the Sigma
service department (see chapter 7.3 - "Service contact").
→
3.7.5
Imbalance monitoring system
A dialog box may pop up or emit a sound signal in order to indicate that the
centrifuge is in the inadmissible imbalance range. If the rotor is loaded
unevenly, the drive will be switched off in the acceleration phase or during
the run.
→
3.7.6
Temperature monitoring system
If the temperature inside the rotor chamber rises above +50°C, the drive
system will be switched off automatically. The centrifuge cannot be
restarted until it has cooled.
→
3.7.7
Rotor monitoring system
When a rotor number and, if applicable, a bucket number are selected, the
computer will automatically check whether the entered speed or the
entered gravitational field are permissible for the selected rotor.