Refrigerated Centrifuge with Heating System Sigma 3-18KHS
6 Using the centrifuge
46 / 89
Version 08/2012, Rev. 1.21 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0701202
Pos : 138 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ Spinc ontrol S (Z ent7)/ 060- 0030-0020-0070 M enü Set up @ 27\mod_1405320225479_68. docx @ 193361 @ 4 @ 1
→
6.3.2.8
Setup menu
The "Setup" menu is symbolised by the "
" symbol on the menu bar. It is
used to perform basic settings concerning the control system of the
centrifuge. It enables the optimum adaptation of the centrifuge to its specific
area of application.
Fig. 20: Menu "Setup"
Function
Open lid after run
The automatic lid opening function ensures that the lid opens when the
rotor has stopped.
CAUTION
When the lid is open, the cooling is not active. The sample may warm up!
Program rotation
See chapter 6.3.3.5 - "Automatic program rotation".
Runtime as of set speed
If this function is active, the runtime will not be measured until the set speed
is reached. In the "Standard" menu, the symbol "
!" will be displayed in the
runtime field.
NOTE
In the program mode, this function can be saved separately for every
program. In this case, the symbol "
!" will be displayed under the runtime
of the detailed program description.